g.Nautilus RESEARCH is a wearable EEG headset for research applications only. This means that it’s less expensive and intended to be used for neuroscience research only. It comes with flexible cables to configure the EEG electrode positions as you wish. A hybrid (dry and wet) EEG electrode version and a gel EEG electrode version with 8/16/32/64 EEG channels are available for any kind of neurological and neuroscience research.
PRODUCT HIGHLIGHTS
Electrode strands can be plugged into the headset |
Strand with g.SAHARA hybrid EEG electrodes for dry and gel recordings |
Strand with g.SCARABEO EEG electrodes for wet recordings |
Strand with safety sockets to connect g.SCARABEO, g.SAHARA hybrid or g.Sensors |
Easy cleaning of the electrode strands; fits ideally for medical applications |
8/16/19/32 channels wearable EEG with 3-axis accelerometer |
24-bit accuracy at 500 Hz sampling rate |
2.4 GHz digital transmission, range: 10 meters indoor |
10 hours continuous recording and 2-3 hours charging |
Internal impedance check with for g.SCARABEO electrodes |
A new benchmark in usability |
Full software environment |
CE certified and FDA cleared medical EEG headset |
TECHNICAL SPECIFICATIONS
Weight | < 110 g without electrode grid |
Size | 78 (L) x 94 (W) x 26 (H) mm |
Color | BLACK |
Sensitivity | ±2.25 V, ±1.125 V, ±750 mV, ±562,5 mV, ±375 mV, ±187.5 mV (software selectable) |
Interface | Wireless 2.4 GHz ISM band |
Digital inputs | 8 digital trigger inputs at Base Station |
Supply | Built-in lithium ion battery, runtime > 10 h with 32 channels, inductive charging according to the QI standard of the Wireless Power Consortium |
Amplifier type | Real DC coupled |
32 × ADC | 24 Bit (1.024 MHz internal sampling per channel) |
Noise level | < 0.6 µV RMS between 1 and 30 Hz (at highest input sensitivity) |
Input channels | 32 mono-polar / 16 bi-polar channels with GND and REF (software selectable) |
Input impedance | DC > 100 MOhm |
Applied part | BF |
Safety class | II |
Certification and Standards | CE certified and FDA cleared medical device IEC60601-1 3rd, IEC60601-1-2, IEC62304, IEC60601-2-26, ISO 14971 10993-1, 62366, 2010-2012 IEEE Recommended Practice for Neurofeedback Systems |